Show simple item record

dc.contributor.authorSamila, Andriy
dc.contributor.authorHotra, Oleksandra
dc.contributor.authorSafronov, Igor
dc.date.accessioned2021-06-16T10:55:50Z
dc.date.available2021-06-16T10:55:50Z
dc.date.issued2020-11-12
dc.identifier.citationSamila A., Safronov I., Hotra O. Structural and functional synthesis of the continuous wave NQR temperature sensor with increased conversion linearity. Solid State Nuclear Magnetic Resonance. 2020. Vol. 110. Р. 101700-1–9.
dc.identifier.issn0926-2040
dc.identifier.urihttps://archer.chnu.edu.ua/xmlui/handle/123456789/62
dc.description.abstractThe paper describes development of the detailed structure and circuit diagrams of the continuous wave NQR temperature sensor with increased conversion linearity. It is experimentally established that at amplitude modulation of 40% and change of input voltage in the range of 20–1000 mV, the circuit of a symmetric marginal oscillator with a linear active demodulator provides better linearity of transfer characteristic than the circuits of asymmetric marginal oscillators with JFET or diode detectors. As a thermometric substance of the proposed NQR sensor, copper oxide Cu2O was used, which is characterized by a strong temperature dependence of the resonance frequency of 63Cu NQR. In contrast to 35Cl NQR in KClO3, for cuprous oxide the temperature dependence of 63Сu NQR frequency in the frequency range 26.621–25.658 MHz is linear in the temperature range 100–390 K. It is experimentally confirmed that the use of a low mass sample (less than 200 mg) as a thermometric substance of the proposed NQR sensor is quite sufficient for successfully observation of the resonance line at the SNR equal to 9.1 dB.uk_UA
dc.description.sponsorshipThis work was supported by the Ministry of Education and Science of Ukraine (grant No 18.801, state registration number 0120U101249).uk_UA
dc.language.isoenuk_UA
dc.publisherElsevieruk_UA
dc.relation.ispartofseriesSolid State Nuclear Magnetic Resonance;110
dc.subjectnuclear quadrupole resonanceuk_UA
dc.subjectmarginal oscillator
dc.subjectamplitude demodulation
dc.subjecttemperature sensor
dc.titleStructural and functional synthesis of the continuous wave NQR temperature sensor with increased conversion linearityuk_UA